Need assistance? Call us: 1-800-663-4209
PX309
General Purpose, Stainless Steel Pressure Transducers
Models In Stock
- 1, 2 & 5 psi Low Pressure Ranges!!
- All Stainless Steel Construction
- Gage, Absolute, Compound Gage or Vacuum Pressure
- Rugged Solid State Design
PX309
From
$
313.50
Models In Stock
Product Overview
- Accuracy Class 0.1% to 0.25%
- Additional Features 1 None
- Bandwidth DC to 1 kHz
- Enclosure Material 316 Stainless Steel
- Enclosure Ratings IP65
- Insulation Resistance 1 MΩ @ 25 Vdc
- Pressure Cycles 10 million, minimum
- Measurement Pressure, Min 0 psi
- Pressure Note Pressure Overload: -15 to 50 psig and 0 to 1000 psia: 3 times rated pressure or 20 psi whichever is greater, 100 to 10,000 psig: 2 times rated pressure
- Process Connection Size 1/4 in
- Process Connection Type NPT Male
- Response Time <1 ms
- Series ID PX309
- Shock 50 g, 11 ms half-sine
- Shunt Calibration Value In vertical direction with fitting down, compound ranges are calibrated for positive pressure output only
- Stability ± 0.25%
- Stability Note 1-Year Stability: <0.25% FS
- Supply Voltage 9 to 30 Vdc
- Vibration ± 20 g
- Weight 154 g
OMEGA’s PX309 Series Stainless Steel Pressure transducers below 100 psi use a high accuracy silicon sensor protected by an oil-filled stainless steel diaphragm. Units 100 psi and above use silicon strain gauges molecularly bonded to the stainless steel diaphragm.
We provide a complete range of services—from product inception, through design and prototypes, to manufacturing and testing. Our application engineers work closely with our customers to customize, design or create entirely new products. Call us—whether you’re an OEM, manufacturer, or end user.
The PX309 is also available in Metric configurations; please see the PXM309 for complete details, or contact our Pressure department for more information.
We provide a complete range of services—from product inception, through design and prototypes, to manufacturing and testing. Our application engineers work closely with our customers to customize, design or create entirely new products. Call us—whether you’re an OEM, manufacturer, or end user.
The PX309 is also available in Metric configurations; please see the PXM309 for complete details, or contact our Pressure department for more information.
PDFs & Manuals
Accessories
accessories
cables
meters/controllers
1/8 DIN Thermocouple and RTD Panel Meter with Large, Selectable Color Display
$535.94 / each
0 In Stock
1/8 DIN Thermocouple and RTD Panel Meter with Large, Selectable Color Display
$535.94 / each
0 In Stock
Show Ratings & Reviews
Rated 1 out of
5
by
Eric M from
Slow, unresponsive ordering process
The entire process of buying to delivery was slow and poorly communicated. I had submitted an urgent order and it took way to long to receive order confirmation as well as unacceptably long for the product to ship. Please make drastic improvements to your order processing process and take actions like actively reviewing the current process for inefficiencies and ways to improve.
Date published: 2025-12-26
Rated 5 out of
5
by
NDPIII from
Solid product
Durable , stable readings in propeller dyno environment .
Date published: 2025-12-10
Rated 5 out of
5
by
Alex I from
No hassle
No hassle, I got the items I ordered. I had a tight timeline.
Date published: 2025-12-08
Rated 5 out of
5
by
NDPIII from
Great product for the price.
We use these in our dyno testing , very durable and reliable .
Date published: 2025-10-10
Rated 5 out of
5
by
Lee J from
Dependable and performs well
Dependable and performs as needed. Trusted quality and no issues so far.
Date published: 2025-10-09
Rated 5 out of
5
by
Henmik67 from
Right part for the right price
Very compact and reliable part, reasonably priced.
Date published: 2025-08-18
Rated 5 out of
5
by
USCshan from
Great and reliable product for Universities
Reliable pressure transducer that works great and is easy to install.
Date published: 2024-11-19
Rated 5 out of
5
by
Tom Dineen from
0psi to 5,000psi transducer
These work great on all of our drill rigs. They're heavy duty enough to work in our harsh environment and still read like new.
Date published: 2024-11-18
Rated 5 out of
5
by
jdakgl from
Transducer works great!
This transducer works great! I have quite of few of these and have always been pleased.
Date published: 2024-11-18
Rated 5 out of
5
by
Anonymous from
Dependable, reliable, and easy to use.
Dependable, reliable, and easy to use. We use on engine Testcells.
Date published: 2024-11-14
Rated 5 out of
5
by
Kpellet from
Availability is great!
Product is always available when I need to place an order. This is super important to us when we have continued issues going on at our facility.
Date published: 2024-11-14
Rated 5 out of
5
by
Sourav Bhunia from
A robust and accurate pressure sensor
The PX309 is robust and accurate pressure sensor. It will be helpful to note in the instructions that the black wire is the common ground for BOTH the power and signal.
Date published: 2024-10-30
Rated 5 out of
5
by
Anonymous from
Good for Experienced Professional
Very professional experience. Good customer service.
Date published: 2024-10-21
Rated 5 out of
5
by
KM713 from
Buy the USB version.
USB version is very easy to use. We bought the ones that go with a DAQ first, had tons of trouble because the DAQ wasn't putting out the right voltage for the sensor.
Date published: 2024-10-14
Rated 5 out of
5
by
Keylin F from
Good
Very simple purchasing process. Solid product as well
Date published: 2024-10-09
Rated 5 out of
5
by
Jeff Scott from
Principal Mechanical Engineer
The gauge works great. It has excellent frequency response and accuracy
Date published: 2024-10-06
Rated 5 out of
5
by
Alex B from
Great sensor
Very good range, easy to wire, very well built and durable-feeling. Came with a NIST certificate, will definitely order more in the future.
Date published: 2024-09-23
Rated 5 out of
5
by
L DiMambro from
Transducer
Product exactly as described. Fast shipping and smooth transaction.
Date published: 2024-09-17
Rated 5 out of
5
by
ClayS from
Right part
They were very helpful in getting the right part ordered
Date published: 2024-08-21
Rated 5 out of
5
by
PeteS from
Good packaging
Great Pricing, packaged well. Have not installed yet, but a quality looking product.
Date published: 2024-07-26
Rated 5 out of
5
by
Randall from
Reliable Quality Sensor
Great item. Easy to set up and use. Used in compressed air and boiler system.
Date published: 2024-07-12
Rated 5 out of
5
by
Justin Dubik from
Good experience using this for experiments
Easy to use and incorporate into our setup for research
Date published: 2024-07-09
Rated 5 out of
5
by
REM_WCEC from
also no surprises
Works as expected and easy to integrate with our NI DAQ system
Date published: 2024-06-04
Rated 5 out of
5
by
REM_WCEC from
no surprises.
Works as expected and, easy set up with our NI DAQ monitoring water pressures for a multi-function heatpump-waterheater.
Date published: 2024-06-04
Rated 5 out of
5
by
Anonymous from
Great Low-Pressure Transducer
Works great, performs exactly as expected for my application.
Date published: 2024-05-28
Rated 5 out of
5
by
Anonymous from
Good quality
We have been using this transducer for more than 10 years on our instruments, good quality!!!
Date published: 2024-05-02
Rated 4 out of
5
by
Terry H from
quality product
This is a quality product for a low-pressure PT. This is a second time purchase though. The original lasted about 4-5 years so I can't complain too much.
Date published: 2024-04-17
Rated 5 out of
5
by
R and D Testing from
Great product
Works as expected, delivery was quick minimal impact to operations
Date published: 2024-03-13
Rated 5 out of
5
by
Anonymous from
Fast Fast Fast Delivery
Item as described, fast delivery, Omega always delivers quality products.
Date published: 2024-02-19
Rated 5 out of
5
by
MikeA from
No issues
Works great second one we bought it works great for what we do
Date published: 2024-02-18
What is the inside diameter and depth of the sensing port within the Male 1/4 NPT port on model PX309-100A10V ?
Thanks for your inquiry.
Would only have available a approximation.
1/8 of an inch The depth would fluctuate
Hope this was helpful.
Date published: 2021-04-30
Will PX309-100G5V work with Both Liquid and gas environments?
Hello, Thank you for contacting Omega Engineering. The PX309 is compatible with both gases and liquids. You will however have to make sure your media is compatible with the wetted parts.
Date published: 2023-02-22
What is the input impedance and output impedance of the PX309?
We do not provide the input/output impedance for the PX309 0 to 5VDC output units. The impedance can widely vary due to the internal components used in the each sensor.
Date published: 2021-03-04
What is the difference between a compound gauge and absolute gauge?
The compound gauge are measuring a vacuum or positive pressure relative to the atmospheric pressure. The unit has a vent to atmosphere for the reference. The absolute gauge units measure pressure relative to an internal sealed vacuum reference.
Date published: 2023-03-14
There are 4 output options"4-20mA, 0to5V,0to10V, and 'mV output'. What I want is a real time dynamic output of the instantaneous pressure level. I know this much:4-20mA does NOT do what I want. Do any of the others match my description?
The response time of the mV/V and DC Voltage output PX309 transducers is approximately 1 millisecond. While the response is fairly quick these units are designed for static pressure measurement. There are dynamic pressure sensors that are of piezo-electric design the have a response in the microsecond range. Omega does not offer this type of sensor
Date published: 2023-02-11
Are the sensors IO link compatible
Thank you for your inquiry, this unit is not IO link compatible
Date published: 2022-09-09
The O'ring is necessary to install PX329-100G5V? What is the torque rage to install the PX329-100G5V?
There is no O-ring used on a NPT fitting. You would typically use teflon tape (plumber's tape wrapped around the threads before you install it into the mating female thread connection. You do not torque a npt fitting. You would hand tighten the connection and the use a wrench to tighten it just enough so it doesn't leak
Date published: 2024-03-11
Can the PX309 be installed onto a hydraulic system? Can it be used with hydraulic oils to take readings without damage? Thank you
The PX309 is constructed with stainless steel wetted materials, which would be compatible with hydraulic oils.
Date published: 2021-05-26
where is the atmospheric reference port located? is it insdie the electrical harness?
All PX309 (gage and absolute) use a vented
cable. Gage PX309 sensors up to 500 PSIG must
be terminated in a clean dry area that is open to
atmosphere for correct operation. All other sensors
(> 500 psig and all psia ranges) must be terminated
in a dry area but do not need to be vented to
atmosphere. These cables should be sealed (e.g.
RTV) to protect the sensor from moisture getting in
to the sensor through the cable.
All PX319 are vented through a hole in the
connector and through the mating connector. A
vented cable must be used for low pressure gage
sensors up to 500 PSIG and must be terminated
in a clean dry area that is open to atmosphere
for correct operation. The gasket for the DIN
mating connector has a special stiffener (included)
between the mating surfaces and is used to
maintain the IP65 rating (Spare Gaskets are Part
number P002612). Search cable assembly
CA-319 with a “WP” on the end for gage pressure
under 500 psi. All other PX319 sensors can use
standard cable without the “WP” but must be
terminated in a dry location or sealed to prevent
water from getting into the cable/sensor.
All PX329 are vented through Pin F of the
twist-lock connector. Gage PX329 up to 500 PSIG
must be vented for correct operation and must use the
connector PT06V-10-6S-R which has no socket on pin
F. Cables for these sensors must use a vented cable,
be sealed at the back of the mating connector and be
terminated in a dry clean area open to the atmosphere.
Search cable assembly CA-329 with “WP” for gage
pressure under 500 psi. All other PX329 can use the
standard PT06F10-6S-R connector or cables (See
cable assembly CA-329 without “WP” ending).
Date published: 2024-11-25
What is the supply current on the PX309 whose output is 100mV?
We cannot provide a fixed value for the current draw of the 100mV output version of the PX309 Series. The impedance can widely vary due to the internal components used in the each sensor. The current draw however would be extremely low.
Date published: 2021-05-04
I am interested in the PX309-015GI pressure transducer. In our application, the transducer will be sitting at a slight vacuum (~1 PSI) when the system we are measuring is not running. Is this an issue for maintaining calibration?
Thank you for your inquiry, 1 PSI vacuum would not be an issue the unit will maintaining calibration
Date published: 2023-02-06
FOr the cable model, what is the WG wire size? What jacket material?
Thank you for your inquire.
The nominal size is 26 awg. Jacket is pvc
Date published: 2022-05-17
Do the PX359 series sensors keep their IP65 rating when paired with Omega's M12 connectors, such as item M12C-PUR-4-S-F-2?
No, the sensor will maintain its IP65 rating, but the M12 connectors and cable will not meet that rating.
Date published: 2021-07-07
What is the accuracy and repeatability for PX359-100GI and PX319-300G5V? I see conflicting values in the product summary, common specificiations, and the datasheet.
Combined accuracy of the PX309 Series is 0.25% BSL. This is a root,sum, square calculation of the linearity, hysteresis & repeatability values for this sensor.
Date published: 2020-04-08
Is the PX359 compatible with Helium?
The wetted parts of the PX309/319/329/359 pressure transducers all stainless so media compatibility would not be an issue. The only possible issue the be getting a good seal on the 1/4-18 NPT thread on the process connection when using it with helium. Normally you would use teflon plumber's tape to seal the thread connection
Date published: 2023-03-27
Can the sensor PX213 be used in oxygen enriched environment, e.g. 100% O2 concentration. The question is whether the O2 will be able to enter the electronic part of the sensor or it will be stopped by membrane and if materials are appropriate
Hello, The PX213 has been discontinued for more than a decade. Please note that the PX213 was not oxygen cleaned or oxygen rated, typically for Oxygen use you will need a sensor that is oxygen cleaned to ensure there is no oil fill/ contaminants that will affect your system. If you have more questions feel free to email us at pressure@dwyeromega.com
Date published: 2023-10-25
Are FPW-15 or PSR-24 recommended as excitation sources for PX3093KG5V transmitter?
Yes, The PX309 will run on any DC voltage between 9 and 30 VDC so either supply will work.
Date published: 2020-10-02
Is there CAD for this model available for download?
Thank you for your inquiry, Unfortunately, we are unable to provide a CAD model for this unit. We do offer the dimensional drawing on the Datasheet.
Date published: 2022-12-21
Good Day Would the PX309 be a recommended sensor to measure Natural Gas pressure at 0-15 Inches WG ?
The PX309 Series transducers do not have any explosion proof ratings. We suggest you look at the PX835-005GI has a possible alternative. It has an ATEX rating.
Date published: 2023-06-01
Is it possible to connect multiple pressure transmitters with each of there readout meters into one PC? How to do the cabling and wiring?
Assuming that you would be using the 4 to 20mA output version on the PX309 you can simply break the connect of the PX309 and wire the input to the data acquisition you are using with your pc in series with the 4 to 20mA signal from the sensor. If the data acquisition system is designed for a DC voltage input you will need to use a shunt resistor to convert the current signal to voltage.
Date published: 2022-08-09
Can this transducer be used with a 24Vdc power supply?
Thank you for contacting Omega Engineering with your question. The PX309 series transducers are available several different output types and not all of them can operate on 24 Vdc. If the model in questions has a 0-5 Vdc, 0-10 Vdc, or 4-20 mA output then yes, they can operate on 24 volts. If it is one with a millivolt output, then the excitation voltage would be 5 or 10 Vdc depending on the pressure range. Please feel free to contact us again if you have any other questions.
Date published: 2019-11-19
For the absolute pressure gauge, where the backside of the diaphragm is vented to?
Thank you for your inquiry, on a absolute pressure sensor the backside of the diaphragm is sealed.
Date published: 2021-12-03
Can the sensor PX309 be used in oxygen enriched environment, e.g. 100% O2 concentration. The question is whether the O2 will be able to enter the electronic part of the sensor or it will be stopped by membrane and if materials are appropriate
Thank you for this inquiry
Due to the construction of our model PX309 series sensors we would not recommend that the sensor be used in oxygen enriched environment.
Date published: 2021-09-17
Can the 50 PSIa model read water pressure? or air only?
Thank you for your question?
This is a all stainless steel unit.
It can handle liquids or gas that are compatible with this material.
Date published: 2019-07-22
What is the Ambient Storage temperature of this device? Can it be stored at -40 to 80C and returned to nominal temperature range without degradation or damage?
Thank you for your inquiry. The Ambient Storage temperature of the PX309 series transducers is between 50 and 80 F.
Date published: 2023-10-11
If I need to measure pressures at 0.05 psia, what pressure transducer would you recommend? Note that I need to make pressure measurements in a vacuum chamber.
thank you for your inquiry. The smallest range and most accurate absolute pressure pressure range unit we offer is the PX409-005A10V. This has an accuracy of 0.08%. We do not offer a 0 to 1 PSIA range in the PX309 or PX409 Series sensors.
Date published: 2020-11-20
For the Gauge Type pressure sensor, does it relative the current atmospheric pressure, or a preset one? Thank you! Best, Renjing
Thank you for your inquiry, on a PSIG sensor the backside of the diaphragm is vented to atmospheric pressure.
Date published: 2021-12-03
Does the PX309-3KGI Model Option come with NIST certificate?
Thank you for contacting Omega Engineering, All our PX309 models will come with a 5 point NIST traceable calibration certificate.
Date published: 2023-02-20
Does the twist lock connector version come with the mating connector?
Thank you for reaching out to Omega with your question. The mating connector is not included and will need to be purchased separately. If you need any further assistance, please reach out to our Tech Support.
Date published: 2020-10-01
Is the PX309-150GI compatible to read liquid Anhydrous Ammonia (NH3) ??
The wetted parts of the PX309-150GI are all 17-4 PH Stainless Steel. Corrosion Resistant Charts indicate it has an excellent "A" rating for use with Anhydrous Ammonia.
Date published: 2020-08-08